Ultra Shield Refrigerated Magnet Technology
Engineering expertise in superconducting magnet design

- 7T USR magnet
Modern magnets are available with shielding of the external magnetic field. This is achieved by use of a second superconducting coil which compensates the magnetic field outside the magnet. Bruker’s UltraShield technology drastically reduces the stray field to close to the magnet and thus strongly reduces the magnetic field to which the operator is exposed. All Bruker USR magnets include refrigeration technology for long helium hold time and minimum number of maintenance intervals.
more on USR magnet technology
The choice of magnet is central in defining an MR system. Its specification and characteristics are of major importance in determining the overall system performance. The optimal combination of magnet geometry and magnetic field strength is critically determined by the specific studies and applications of the end user as well as by siting considerations. Bruker BioSpin MRI manufactures a range of horizontal superconducting magnet systems with bore diameters ranging from 160 mm up to 400 mm and nominal magnetic field strengths ranging from 4.7 Tesla up to 11.7 Tesla.
Horizontal USR Magnets with Active Shielding and Refrigeration
| Magnet | 47/40 | 70/20 | 70/30 | 94/20 | 94/30 | 117/16 |
| Magnetic Field (T) | 4.7 | 7.0 | 7.0 | 9.4 | 9.4 | 11.7 |
| Bore diameter (cm) | 40 | 20 | 30 | 20 | 30 | 16 |
| Free access for imaging (mm) | 198 | 86 | 154 | 86 | 154 | 72 |
| Flange to field center (mm) | 745 | 631 | 745 | 745 | 1004 | 692 |
| Max. drift rate (ppm/h) | 0.05 | 0.05 | 0.05 | 0.05 | 0.05 | 0.05 |
| 5 Gauss radial/axial (m) | 2.0/3.0 | 1.5/1.5 | 2.0/3.0 | 2.0/3.0 | 2.3/3.3 | 1.7/2.8 |
| Homogeneous volume Peak-to-Peak (ppm)/dsv (mm) | +/- 2.5/180 | +/- 5/120 | +/- 5/180 | +/- 5/120 | +/- 5/180 | +/- 2/60 |
| Min. He refill intervall (days) | ZBO* | ZBO* | ZBO* | ZBO* | ZBO* | ZBO* |
| Min. ceiling height (cm) | 285 | 260 | 285 | 285 | 290 | 285 |
| Weight (kg) | 4700 | 2500 | 5000 | 5500 | 11500 | 7000 |